• XE Series

XE Series Semi-micro Balance


Readability: 0.01mg,Capacity: 31g-220g;Brand New Monolithic Sensor

New super monolithic sensor technology - accurate, fast, stable and reliable

Compliance with 21CFR Part 11 - full traceability

Operating system with full touch screen

Automatic calibration and adjustment technology

Automatic leveling function

Built-in multi-applications, print output, & data output meet GLP/GMP requirements

Parameter
Introduction
Parameters
Compliance
Options
  • Readability (d)

    0.01mg

  • Max capacity

    31g51g81g120g220g

  • Minimum initial weighing

    1mg

  • 21 CFR Part 11

    Applicable

  • Ionizer

    Optional

  • Interface

    USB-CRS232RS485

  • Type of balance

    Semi-micro Balance

  • Application industries

    High-end LaboratoryPharmaceutical Industry

  • Adopt new super monolithic sensor technology - accurate, fast, stable and reliable.
     ♦ Sensitivity: unparalleled sensor design and processing technology ensure the sensitivity of micro-weighting results
     ♦ Precision: Particular super monolithic sensor technology ensures the accuracy, stability and consistency of weighing results.
     ♦ High-Speed: The sensor technology of horizontal bar, combined with excellent circuit design and software algorithm, meets the requirements of rapid weighing in the laboratory
    Compliance with 21CFR Part 11 - full traceability
     ♦ Designed in accordance with FDA data integrity principles to achieve attributable, legible, contemporaneous, original, and accurate data (ALCOA).
     ♦ User administration: Three-level user administration through access control
     ♦ Electronic Signature: The user''s electronic signature can be used to sign the final report of the weighing operation.
     ♦ Audit Trail and Weighing Log: An audit trail is a tamper-proof electronic log file with a timestamp that allows the reconstruction of events related to the creation, modification, and/or deletion of records.
     ♦ Data transmission security; Electronic records should be protected from any tampering.
     ♦ Data output meets GLP/GMP requirements
    Operating system with full touch screen
     ♦ Operation requirement is concise and clear, so customers can conveniently operate the balance and directly read the results on the screen to have a wonderful experience. At the same time, it is necessary to consider the special requirements of users in special industries, such as non-contact operation, etc.
    Adopt automatic calibration and adjustment technology
     ♦ isoCAL automatic internal calibration and adjustment function, automatically triggered by temperature and time, fully ensure the accuracy of weighing results.
    Adopt automatic door technology (optional)
     ♦ The technology of automatically opening and closing door is not only to improve customer experience satisfaction, but also necessary to consider the requirements of non-contact weighing in special industries, so as to achieve the goal of not contacting the balance during the weighing operation. Besides, the balance has a function of memorizing door position.
    Fully automatic electrostatic removal technology (optional)
     ♦ Built-in ionizer can automatically eliminate the effect of static electricity on weighing
    Built-in multi-applications
     ♦ Built-in multi-applications, print output, & data output meet GLP/GMP requirements
    Standard communication interface
     ♦ Automatically identify the printer of D&T
     ♦ Standard RS232, easy to connect with peripherals
     ♦ PC connection function not only is easy to connect to PC, but also can directly transfer weighing data to spreadsheets (Excel)
    Standard USB-C interface for data storage and export functions

    Model

    XE305S

    XE505S

    XE805S

    XE1205S

    XE2205S

    Max capacity(Max)

    31g

    51g

    81g

    120

    220

    Readability(d)

    0.01mg

    0.01mg

    0.01mg

    0.01mg

    0.01mg

    Repeatability(5% load)

    Tolerance

    0.01mg

    0.01mg

    0.01mg

    0.015mg

    0.02mg

    Typical value

    0.02mg

    0.02mg

    0.02mg

    0.025mg

    0.03mg

    Repeatability(approx. 100% load)

    Tolerance

    0.015mg

    0.015mg

    0.015mg

    0.02mg

    0.025mg

    Typical value

    0.02mg

    0.02mg

    0.03mg

    0.03mg

    0.05mg

    Linearity deviation

    Tolerance

    0.015mg

    0.015mg

    0.015mg

    0.02mg

    0.03mg

    Typical value

    0.02mg

    0.02mg

    0.02mg

    0.03mg

    0.03mg

    Off-center load deviation ( positions according to OIML R76)

    Test weight

    20g

    50g

    50g

    100g

    200g

    Tolerance

    0.02mg

    0.02mg

    0.03mg

    0.04mg

    0.04mg

    Typical value

    0.05mg

    0.05mg

    0.05mg

    0.06mg

    0.08mg

    Sensitivity drift

    (+10℃~+30℃)

    8 ppm/K

    Tare max capacity:<100% max capacity

    Accuracy class

    Ⅰ( according to Directive 2014 | 31 | EU)

    Verification scale interval(e)

    ( according to Directive 2014 | 31 | EU)

    0.1mg

    0.1mg

    0.1mg

    0.1mg

    0.1mg

    Minimum load

    ( according to Directive 2014 | 31 | EU)

    1mg

    1mg

    1mg

    1mg

    1mg

    Minimum initial weighing :according to USP(United States Pharmacopeia) Chap. 41

    Optimum minimum initial weighing

    0.82mg

    0.82mg

    0.82mg

    0.82mg

    0.82mg

    Typical minimum initial weighing

    1.3mg

    1.3mg

    1.3mg

    1.3mg

    1.3mg

    Typical stabilization time

    3s

    3s

    3s

    3s

    5s

    Typical measurement time

    5s

    5s

    5s

    5s

    8s

    Recommended calibration weight

    External calibrated test weight

    20g

    50g

    50g

    100g

    200g

    Accuracy class

     (according to OIML R111-1)

    E1

    E1

    E1

    E1

    E1

    Dimensions

    L x W x H

    240mm×430mm×380mm

    Weighing pan

    85mm×85mm

    Weighing chamber height

    248mm

    Net weight

    approx. 9.86kg

    Ambient conditions

    Operation temperature

    18℃~23℃

    Relative humidity

    50%RH~75%RH

    Warm-up temperature fluctuation

    ≤1°C/h

    Power supply

    100V-240VAC 50HZ/60HZ 1.0A-0.5A

    ♦ GB/T26497-2022 Electronic Balances

    ♦ JB/T9511-1999 Balance-packing Requirements

    ♦ OIML R76-2(2007E) International Recommendation: Non-automatic weighing instruments

    1.Automatic door technology

    2.Automatic electrostatic removal technology

    3.Second screen

    4.Pharmaceutical software package

    5.External draft shield

    6.Gesture sensor

    7.Portable micro printer

    All rights reserved. ©2023-2024 Tianjin Deante Sensor Technology Co., LTD     |     津ICP备19010295号     |