Readability (d)
1μg (0.001mg)
Max capacity
2.1g5.1g31g
Minimum initial weighing
1mg
21 CFR Part 11
Applicable
Ionizer
Optional
Interface
USB-CRS232RS485
Type of balance
Ultra-micro Balance
Application industries
High-end LaboratoryPharmaceutical Industry
|
Model |
XE2.6S |
XE5.6S |
XE31.2.6P |
XE31.5.6P |
XE31.6S |
|||
|
Max capacity(Max) |
2.1g |
5.1g |
2.1g/31g |
5.1g/31g |
31g |
|||
|
Readability(d) |
1μg |
1μg |
1μg/10μg |
1μg/10μg |
1μg |
|||
|
Repeatability(5% load) |
||||||||
|
Tolerance |
2μg |
2μg |
2μg/10μg |
2μg/10μg |
2μg |
|||
|
Typical value |
2.5μg |
2.5μg |
2.5μg/20μg |
2.5μg/20μg |
2.5μg |
|||
|
Repeatability(approx. 100% load) |
||||||||
|
Tolerance |
1.5μg |
2.5μg |
2.5μg/15μg |
2.5μg/15μg |
2.5μg |
|||
|
Typical value |
3μg |
4μg |
3μg/20μg |
3μg/20μg |
5μg |
|||
|
Linearity deviation |
||||||||
|
Tolerance |
2μg |
2μg |
2μg/10μg |
2μg/10μg |
2μg |
|||
|
Typical value |
2.5μg |
2.5μg |
2.5μg/20μg |
2.5μg/20μg |
5μg |
|||
|
Off-center load deviation ( positions according to OIML R76) |
||||||||
|
Test weight |
1g |
2g |
2g/20g |
2g/20g |
20g |
|||
|
Tolerance |
2μg |
2μg |
2μg/10μg |
2μg/10μg |
2μg |
|||
|
Typical value |
2.5μg |
2.5μg |
2.5μg/20μg |
2.5μg/20μg |
5μg |
|||
|
Sensitivity drift (+10℃~+30℃) |
5 ppm/K |
5 ppm/K |
5 ppm/K |
5 ppm/K |
5 ppm/K |
|||
|
Tare max capacity:<100% max capacity |
||||||||
|
Accuracy class |
Ⅰ( according to Directive 2014 | 31 | EU) |
|||||||
|
Verification scale interval(e) (according to Directive 2014 | 31 | EU) |
10μg |
10μg |
10μg/100μg |
10μg/100μg |
10μg |
|||
|
Minimum load (according to Directive 2014 | 31 | EU) |
1mg |
1mg |
1mg |
1mg |
1mg |
|||
|
Minimum initial weighing :according to USP(United States Pharmacopeia) Chap. 41 |
||||||||
|
Optimum minimum initial weighing |
0.82mg |
0.82mg |
0.82mg |
0.82mg |
0.82mg |
|||
|
Typical minimum initial weighing |
1.3mg |
1.3mg |
1.3mg |
1.3mg |
1.3mg |
|||
|
Typical stabilization time |
6s |
6s |
6s |
6s |
8s |
|||
|
Typical measurement time |
8s |
8s |
8s |
8s |
12s |
|||
|
Recommended calibration weight |
||||||||
|
External calibrated test weight |
2g |
5g |
2g/20g |
5g/20g |
20g |
|||
|
Accuracy class (according to OIML R111-1) |
E1 |
E1 |
E1 |
E1 |
E1 |
|||
|
Dimensions |
||||||||
|
L x W x H |
90mm×295mm×150mm |
|||||||
|
Weighing pan |
Φ30mm |
|||||||
|
Weighing chamber height |
80mm |
|||||||
|
Net weight |
approx. 4.7kg |
|||||||
|
Ambient conditions |
||||||||
|
Operation temperature |
18℃~23℃ |
|||||||
|
Relative humidity |
50%RH~75%RH |
|||||||
|
Warm-up temperature fluctuation |
≤1°C/h |
|||||||
|
Power supply |
100V-240VAC 50HZ/60HZ 1.0A-0.5A |
|||||||
♦ GB/T26497-2022 Electronic Balances
♦ JB/T9511-1999 Balance-packing Requirements
♦ OIML R76-2(2007E) International Recommendation: Non-automatic weighing instruments
1.Automatic door technology
2.Automatic electrostatic removal technology
3.Second screen
4.Pharmaceutical software package
5.External draft shield
6.Gesture sensor
7.Portable micro printer